top of page

The dust measuring device PCE-MPC 20 is a particle counter that measures the particles in the atmosphere. The measurable particle sizes of the dust measuring device are 0.3, 2.5 and 10 μm. In addition to the particle counter function, the dust meter also measures the mass concentrations PM2.5 and PM10. The built-in high-performance pump ensures a sufficient flow rate at the dust measuring device. In addition, the dust meter can also measure the temperature, humidity and dew point.

 

The measured values ​​can be read directly on the display of the dust measuring device. Data storage is also possible with the dust measuring device in order to analyze the measured values ​​after a measuring process. The battery used ensures a mobile application. This allows a mobile continuous operation of about 4 hours. A dust measuring device is used wherever a possible health hazard due to fine dust could occur. These include pollution from incineration, manufacturing, energy production, vehicle engine emissions, and the pollution that the construction industry generates.


- Mass concentration PM2.5 / PM10
- Data memory up to 5,000 data sets
- Flow rate 2.83 l / min
- Display with graphic real-time diagram
- Battery or AC operation possible
- Alarm function

 

 

MPC 20 Manual Download

 

Air Sampler MPC 20 | 微塵分析儀

  • Mass concentration  
    Measurable particle sizes PM2.5 / PM10
    Measuring range mass concentration 0 ... 2000 μg / m³
    Resolution 1 μg / m³
       
    Particle counter  
    Measurable particle sizes 0.3, 2.5 and 10 μm
    Flow rate 2.83 l / min (0.1 ft³)
    Coincidence error < 5% at 2,000,000 particles per ft³
    Counting efficiency 50% at 0.3 μm;
      100% with particle size > 0.45 μm
    Counting modes Cumulative, differential, concentration
       
    Measurement of air temperature and relative humidity
    Measuring range air temperature 0 ... 50°C / 32 ... 122°F
    Measuring range dew point temperature 0 ... 50°C / 32 ... 122°F
    Measuring range relative humidity 0 ... 100%
    Measuring accuracy air temperature 10 ... 40°C / 50 ... 104°F: ± 1.0°C / 1.8°F
      Others: ± 2.0°C / 3.6°F
    Measuring accuracy dew point temperature      10 ... 40°C / 50 ... 104°F: ± 1.0°C / 1.8°F
      Others: ± 2.0°C / 3.6°F
    Measuring accuracy relative humidity 0 ... 20% rel. Humidity: ± 5%
      20 ... 80% rel. Humidity: ± 3.5%
      80 ... 100% rel. Humidity: ± 5%
    Operating conditions 0 ... 50°C / 32 ... 122°F, 10 ... 90% rel. humidity,
      Not condensing
    Storage conditions -10 ... 60°C / 14 ... 140°F, 10 ... 90% rel. humidity,
      Not condensing
    Data storage 5,000 data sets (Micro SD card)
    Display 2.8", 320 x 240 pixels
      Backlit color LCD
    Power supply
    Charger
    Rechargeable battery 1200-mAh, 7.4V
    100 ... 240 VAC, 50 or 60 Hz
    Battery Life About 4 hours in continuous operation
    Battery recharge time About 2 hours with AC adapter
    Dimensions 240 x 75 x 57 mm / 9.4 x 3 x 2.2 in
    Weight 570 g / 1.3 lbs
bottom of page