Regular calibration of temperature sensors is crucial, as they are exposed to a variety of mechanical, thermal and chemical stresses that can lead to drift over time. This drift can cause the measured temperature to deviate from the actual temperature. Through regular calibration, the differences between the actual and measured temperature can be determined, making the specific drift behaviour visible. Dry-block calibrators are effective instruments for checking and calibrating a wide range of temperature measuring devices and temperature sensors.

The calibrator / simulator is used to calibrate temperature sensors such as Pt100, Pt1000 and many other temperature sensors. The dry block function offers a practical solution for calibrating temperature sensors in the laboratory and in the field. The use of the correct "transition sleeve" ensures optimum thermal coupling between the dry block and the test item to be calibrated.

The versatility of the calibration insert makes it possible to calibrate a wide range of straight temperature sensors with different lengths and diameters with the calibrator / simulator. This significantly increases the flexibility and applicability of the dry block. Another advantage of the dry block function is that it covers the entire temperature range without having to change the calibration medium. This significantly reduces the effort for the user, as no special precautions need to be taken for different viscosities, flash points or outgassing.

Overall, the calibrator / simulator offers an efficient and user-friendly method for calibrating temperature sensors, both in the laboratory and in field applications.

Calibrator / Simulator PCE-DBC 650

  • - Simple calibration

    - 4 sensor openings 6 ... 12 mm

    - Easily transportable

    - colour touch display

    - adjustable alarm

    - PID settings

    - internal memory

  • Temperature  
    Measuring range   300 °C ... 1200 °C
    Resolution   0.001 °C
    Accuracy   0.1 %
    General technical data  
    Units   °C, °F
    Display type   Touch display
    Display size   4.96 inch
    Storage medium   Internal memory
    Storage capacity   50 data records
    Temperature stability   ≤±0.2 °C/15 min
    Vertical temperature field   Deviation within 10 mm of floor calibration insert is 1 °C
    Horizontal temperature field   ≤±0.25 °C (equipped with thermostat)
    Heating speed   25 ... 100 °C: 10 min
    100 ... 600 °C: 15 min
    600 ... 800 °C: 20 min
    800 ... 1200 °C: 30 min
    Cooling speed   1200 ... 800 °C: 25 min
    800 ... 600 °C: 15 min
    600 ... 300 °C: 60 min
    300 ... 50 °C: 180 min
    Dimensions   Calibration insert:
    4 openings for sensors: 6, 8, 10, 12 mm
    Depth: 135 mm
    Fuse(s)   Fuse 20 A, 250 V
    Menu language   English, Chinese
    Protection class (device)   IP20
    Power supply   110 ... 220 V AC, 45 ... 65 Hz
    Plug type   Protective contact plug
    Weight   7.7 kg
    Dimensions (L x W x H)   295 x 174 x 350 mm
    Operating conditions   0 ... 50 °C , 0 ... 90 % r.H
    Storage conditions   0 ... 50 °C , 0 ... 90 % r.H

     

Copyright© 2024 by Tops Instruments Supplies Co.  graphics, data sheet, photos and videos all rights reserved. 