| PCE-PQC 34EU |
|
| Measuring range |
0.3 ... 25 µm |
| Measuring channel sizes |
factory calibrated at 0.3, 0.5, 1.0, 2.5, 5.0, 10.0 µm |
| |
|
| Counting efficiency |
50% at 0.3 μm, 100% at > 0.45 μm according to JIS |
| Particle mass display |
PM0.5, PM1.0, PM2.5, PM5.0, PM10 and TPM |
| Flow rate |
2.83 l/min (0.1 ft³/min) |
| Random loss |
10% at 30000000 particles / ft³ |
| Battery operation |
10 h |
| Light source |
long-life laser diode |
| Zero count |
<1 count / 5 min (<2 particles/ft³) according to ISO 21501-4 and JIS |
| Counting modes |
automatic, manual, real-time display, cumulative/ |
| |
differential, mass concentration |
| Acoustic alarm |
adjustable, integrated alarm |
| Calibration |
traceable to NIST |
| Display |
4.3" WQVGA color touch display, 480x272 px |
| Printer |
external thermal printer |
| Air intake |
internal pump with automatic |
| |
flow rate control |
| Air outlet |
interner HEPA filter |
| Battery |
replaceable Li-Ion battery |
| Charging time |
ca. 4 h |
| Reports |
ISO 14644-1 |
| |
EU GMP appendix 1 |
| |
FS 209E |
| Configuration |
memory for 50 user-defined configurations |
| Standards |
ISO 21501-4 and JIS B9921 |
| Dimensions |
13.3 x 10.5 x 21 cm |
| Weight |
1.8 kg |
| Memory |
45000 data records (ring memory) consisting of |
| |
particle counting, temperature and humidity, CO2, places and times |
| |
|
| Sample duration |
1 s ... 99 h adjustable |
| Power supply |
110 ... 240 V AC 50/60 Hz |
| Operating conditions |
+5 ... +40 °C |
| |
up to 95% r.H. not condensing |
| Storage conditions |
0 ... +50 °C |
| |
up to 98% r.H. not condensing |
| |
|
| CO2 Sensor |
|
| Measuring range |
0 ... 5000 ppm |
| Resolution |
1 ppm |
| Accuracy |
±1 % of measuring range |
| |
|
| |
|
| Temp/humidity sensor internal |
0 ... +50 °C, 15 ... 90 % r.H. |
| Resolution |
0.5 °C |
| Accuracy |
±0.5 °C, ±2 % r.H. |
| |
|
| Interface |
Ethernet, USB |
| Optional interface |
Wifi 802.11 b/g, RS485 or RS232 |
| Number of measuring channels |
6 |
| Housing material |
stainless steel |