top of page

The surface tester works according to the magnetic inductive (ISO 2178) and the eddy current method (ISO 2360). These methods supported by the surface tester are used for the non-destructive testing of materials. The surface tester is used to measure the thickness of magnetically neutral layers on a magnetic or non-magnetic base material.

 

The surface tester is ideal for reliable layer thickness measurements on site. With the external probe, the layer thickness can be measured quickly and easily, even in hard-to-reach places. The surface tester was developed for non-destructive, fast and precise layer thickness measurements and is easy to use. Measured data can be easily transferred to a PC using a USB data storage device or via WiFi.

 

The surface tester is used in particular in the finishing industry, electroplating, ship and bridge construction, aircraft construction and the machine and chemical industry.

 

CT 100N Manual Download

Coating Thickness Gauge CT 100N

  • - Illuminated touch keyboard
    - Measurement data storage for up to 1260 data
    - One-point and multi-point calibration
    - Single measurement and continuous measurement
    - Data transfer via wifi and USB
    - Two different probes for NFe and Fe

  • Measuring range 0 ... 1250 µm; 0 ... 49.2 mil
    Min. coating thickness of base material 50 µm
    Smallest measuring surface (diameter) 10 mm
    Min. radius of curvature convex 15 mm
    Min. radius of curvature concave 15 mm
    Accuracy ± (1% + 1 µm); ± (1% + 0.04 mil)
    Resolution 0.1 µm (<99.9 µm); 0.004 mil (<3.9 mil)
      1 µm (> 100 µm); 0.04 mil (> 3.9 mil)
    Calibration Single point, multiple point
    Data storage 1260 data
    Measuring modes Single measurements, continuous measurement
    Interfaces Wifi , USB
    Operating conditions 0 ... 50°C / 32 ... 122°F, 20 ... 90% RH non-condensing
    Dimensions 170 x 85 x 35 mm / 6.7 x 3.3 x 1.4 in
    Weight Approx. 335 g / < 1 lb (with battery)