top of page

The particle counters of the PQC 1xEU series measure the concentration of particles such as dust, soot, pollens and many other aerosols in the air. To accurately determine the degree of pollution of the air, the particle counters were developed. Pollution is mainly generated by combustion, material processing, manufacturing, power generation, vehicle engine emissions and the construction industry. With the help of the PQC 1xEU series particle counters the exact amount of dirt particles in the air can be measured.

 

Of greater importance is the degree of pollution of the air with pathogenic particles such as soot, which are released by the industry and especially by diesel vehicles without special filtering in the air. Such dispersion particles are responsible inter alia for reduced visibility, the inhalation of toxic substances and thus reduced labor productivity. It has also been known for some time that the particles also make a not inconsiderable contribution to many medical diseases such as asthma, bronchitis, skin and lung diseases. The particle counters are designed for easy and quick use. The particle counters of the PCE-PQC 1xEU series operate in different modes (real-time, cumulative, differential, mass concentration, ...) and display the results on the display,


- Internal memory
- Particle sizes up to 75 μm
- Color display
- Handy
- 6 measuring channels
- ISO 14644-1, EU GMP Annex I, FS 209E
- Mass concentration
- Ethernet, USB
Incl. calibration certificate traceable to NIST ISO 21501-4 and JIS B9921

 

#Dust Monitor #PM 1 #無塵間 #醫療級別

Dust Monitor PQC 15EU | 醫療級別塵粒分析儀

  • Measuring range 0.3 ... 75 μm
    Measuring channel sizes Factory calibrated at 0.5, 1.0, 2.5, 5.0, 10.0, 30.0 μm
    Counting efficiency 50% at 0.5 μm 100% at > 0.75 μm according to JIS
       
    Flow 2.83 l / min (0.1 ft3 / min)
    Random loss 5% at 4,000,000 particles / ft³
    Battery 10 h
    Light source Long-life laser diode
    Zero count < 1 count / 5 min (< 2 particles / ft³)
      According to ISO 21501-4 and JIS
    Counting modes Automatic, manual, real-time, cumulative /
      Differential, mass concentration
    Alarms 1 ... 9999999 counts, adjustable
    Calibration Traceable to NIST
    Display 4.3" WQVGA color touch display, 480x272 px
    Printer External thermal printer
    Aspiration Internal pump with automatic
      Flow control
    Air outlet Internal HEPA filter
    Battery pack Replaceable Li-Ion battery
    Charging time About 4 hours
    Reports ISO 14644-1
      EU GMP Annex 1
      FS 209E
    Configuration Memory for 50 custom configurations
    Standards ISO 21501-4 and JIS B9921
    Dimensions 25.4 x 12.9 x 11.4 cm
    Weight 1.0 kg
    Storage 45000 data sets (ring memory) consisting of
      Particle count, temp and humidity, locations and times
    Sample locations Up to 1000 locations can be stored
    Samples duration 1 s ... 99 h adjustable
    Power supply 110 ... 240V AC 50/60 Hz
    Operating conditions 5 ... 40°C / 41 ... 104°F
      Up to 95% RH not condensing
    Storage conditions 0 ... 50°C / 32 ... 122°F
      Up to 98% RH not condensing
       
    Interface USB
    Optional interfaces Ethernet, WiFi 802.11 b / g, RS485 or RS232
    Number of measuring channels           6
bottom of page